Regulating AI Adaptation: An Analysis of AI Medical Device UpdatesRegulating AI Adaptation: An Analysis of AI Medical Device Updates
AI summary
The authors systematically analyzed the frequency and nature of updates to FDA-approved AI medical devices, finding that fewer than 2% of device reports were updated through retraining with new data. Nearly a quarter of device reports were updated in the form of new features and marketing claims. A pneumothorax detection model case showed that AUC dropped by up to 0.18 when the model was evaluated at new sites, and retraining with site-specific data recovered up to 0.23 AUC; however, performance at the original site also dropped significantly after retraining, and the authors argue this example challenges the current one-size-fits-all regulatory approval approach.